Abstract

Unsymmetrical benzoN 4 nickel(II) macrocyclic complexes have been explored as an ionophore for preparing PVC-based membrane electrodes selective to SCN −. The resulting electrode exhibited the best working concentration range (1.4×10 −7–0.1 M) with a Nernstian slope of (−59.7±0.9 mV per decade) in the wide pH range of 3.5–10.5. The proposed electrode has a very fast response time (5 s) and high selectivity for thiocyanate over variety of anions. This sensor was successfully applied as an indicator electrode to potentiometric titration of thiocyanate ion with AgNO 3 and for direct determination of the amount of thiocyanate in urine and saliva samples in order to distinguish between smokers and nonsmokers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.