Abstract
We describe a rapid, sensitive procedure for detecting benzodiazepines and tricyclic antidepressants in serum. After the compounds are adsorbed on charcoal and eluted with an organic solvent mixture, the drugs are thin-layer chromatographed and detected by their fluorescence. Because it is so fast and easy (turn-around time of about 1 h), the method can be used for emergency toxicology screening. An analysis requires 1.0 mL of serum, and the limit of detection for each of the drugs is about 0.75 mg/L. The procedure is sensitive enough to allow the use of serum instead of urine for the detection of these drugs, because serum has certain advantages over urine as the sample and better reflects the state of the patient's toxicity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.