Abstract

For the first time, the electrospray ionization efficiency (IE) scales in positive and negative mode are united into a single system enabling direct comparison of IE values across ionization modes. This is made possible by the use of a reference compound that ionizes to a similar extent in both positive and negative modes. Thus, choosing the optimal (i.e., most sensitive) ionization conditions for a given set of analytes is enabled. Ionization efficiencies of 33 compounds ionizing in both modes demonstrate that, contrary to general practice, negative mode allows better sensitivity for 46% of such compounds whereas the positive mode is preferred for only 18%, and for 36%, the results for both modes are comparable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.