Abstract

A micro-sized oxyacetylene flame was generated using a small concentric tube system. Hot gas flow and primary reactive species from the micro-flame were directed toward a developed TLC gel plate to thermally desorb and ionize analytes on the gel surface. The resulting analyte ions subsequently entered the MS inlet for detection. A 1-1.5-mm-wide light-brown line was observed on the TLC plate after the desorption FAPCI/MS (DFAPCI/MS) analysis, revealing that the gel surface withstood a high temperature from the impact of the micro-flame. Volatile and semi-volatile chemical compounds, including amine and amide standards, drugs, and aromatherapy oils, were successfully desorbed, ionized, and detected using this TLC/DFAPCI/MS. The limit of detection of TLC-DFAPCI/MS was determined to be 5ng/spot for dibenzylamine and ethenzamide. TLC/DFAPCI/MS is one of the simplest TLC-MS interfaces showing the advantages such as low costs and an easy set up. The technique is useful for characterizing thermally stable volatile and semi-volatile compounds in a mixture.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.