In this work is presented the fabrication of a thin film membrane as a bio-transducer for aural assistance detection, therefore it will operate at low pressure. The resonant membrane was deposited by PECVD technique at low temperature of deposition T=270°C, using SiH4, GeH4, and Boron gases. The membrane was suspended on a micromachined crystalline silicon frame obtained by wet chemical etching. The a-SiGe:B film presented a resistivity of 2.46×103 (Ω-cm), resistance of 20.8kΩ. Using these experimental data we succeeded in designing a simple structure for sensing low pressure variations. The output voltage of the sensor was measured for a range of pressure from 0 to 3000Pa and at bias voltage of 10V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.