Abstract

The stirring frequency exponent of -1/2 in the theoretical Levich expression appears to apply to PAMPA and Caco-2 assays, where efficient individual-well magnetic stirring ( > 20 RPM) is used. If a single molecule is used as a stirring calibrant, then scaling suggested by Eq. (5) with microtitre plate data may be used. The error in calculating hABL based on unscaled hABLref can be as high as 30%. This is of practical importance in PAMPA, and perhaps cellular assays as well.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.