Abstract

We present results obtained with double-sided, thick (2 mm) silicon strip detectors used as a Compton imager. Reconstructed images and spectra from /sup 137/Cs and /sup 60/Co gamma-ray sources have been produced at room temperature using the multiple Compton technique. Multiple Compton interactions allow the energy and Compton scattering angle to be reconstructed without having to absorb the energy of the incident gamma ray completely. This extends our work on multiple Compton imagers using germanium strip detectors to silicon detectors that operate at higher temperatures. The energy resolution of the detectors for 60 keV gamma rays is 3-4 keV at room temperature and 2.1 keV at -20/spl deg/C. Simulations of the imager have been performed in GEANT4, including Doppler broadening and accidental coincidences, and show agreement with the measured data. Reconstructing the energy spectrum for 662 keV gamma rays that did not deposit their full energy in the instrument shows a peak at 662 keV with a full-width at half-maximum (FWHM) of 27.6 keV. The reconstructed image of the source shows an angular resolution of 3.3/spl deg/ FWHM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.