Resistance measurements of thick platinum films between 4.2 and 300 K are reported. The thick films were prepared by standard screen-printing techniques. The properties of the films make them suitable for low temperature thermometry. In the temperature range 77–300 K experimental data are fitted using a quadratic fit with an error in temperature of less than 0.2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.