Abstract

The thermal stabilization system of the RED-100 liquid-xenon two-phase emission detector for the experiment aimed at detecting the effect of coherent neutrino scattering off xenon nuclei has been tested. The system is based on thermosyphons (closed two-phase heat pipes) that are filled with nitrogen and use liquid nitrogen boiling at atmospheric pressure as a cooler. The system is capable of keeping the working medium of the RED-100 detector at a temperature of 167 K with a precision of <1 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.