Abstract

A thermospray liquid chromatography/mass spectrometry (TSP-LC/MS) method is described for determination of the enantiomeric excess of α-amino acids and α-amino acid amides as their o- phthalaldehyde/N- acetyl- l-cysteine (OPA/NAC) derivatives. The source temperature is an important factor in optimizing the sensitivity of the TSP-LC/MS analysis, whereas the repeller voltage is of minor importance. On-column mass spectra were acquired for the OPA/NAC derivatives of several α-amino acids and α-amino acid amides. For the main fragment ions, mass spectra fragmentation pathways are proposed. The applicability of the method is demonstrated by means of the enantiomeric excess determination of valine in a sample from an enzymatic hydrolysis experiment. Using single ion monitoring, the detection limit of d-valine in the presence of excess l-valine is 10 pmol. The present TSP-LC/MS method is useful for validating the results obtained from LC/UV or LC/fluorescence methods for the enantiomeric excess determination of α-amino acids and α-amino acid amides.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.