Abstract

Chromel–AuFe 0.07% thermocouples are widely used to measure small temperature differences down to liquid-He temperatures. In this work we present the thermopower S of two chromel–AuFe 0.07% thermocouples over an extended temperature range from 1.5 K to 200 K in magnetic fields up to 7 T. For low temperatures T < 4 K we find a sizeable field dependence of S even in small fields. Furthermore, a field of 7 T affects the thermopower considerably up to 200 K. Thus, our data demonstrate a significant field dependence of the thermopower of chromel–AuFe 0.07% thermocouples in an extended temperature and field range compared to previous works.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.