Abstract
The thermometric titration of copper(II) ions with triethylenetetramine was investigated. The apparent enthalpy change of the titration reaction varied over a wide range, even from exothermic to endothermic, depending on the reaction conditions, such as pH of sample and titrant solutions, species of buffer and coexistence of auxiliary reagent. In acetate buffer solution, this reaction was endothermic, and copper(II) ions could be titrated with error less than 0.5% of RSD even in the presence of other heavy metal ions. Furthermore, the sequential titration of copper(II) and zinc(II) ions was possible with error less than 0.7% of RSD when tartrate was added as an auxiliary reagent to get sharp end points. The methods were applied to the analysis of four samples of copper-based alloy. The analytical results agreed closely with those obtained by the JIS (Japanese Industrial Standard) method based on electrogravimetry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.