Abstract

To meet the need for a rugged, easily read thermometer having a flexible probe, a relatively simple instrument has been devised and tested. It consists of a thermistor that rapidly changes in electrical resistance with small changes in temperature, a source of current, and a device to measure the flow of current that is calibrated in such a way as to read temperature directly. It is designed to operate linearly in the range in which experimental and clinical work in hypothermia is involved, 20 C (68 F) to 41 C (105 F). For a discussion of other types of instruments for measuring temperature, the review by Burton1may be consulted. The apparatus described by Weiss and Ripstein,2which is useful for continuous recording, has the disadvantage of depending on a flexible capillary tube that may break or change its characteristics on continued bending. The temperature-sensitive thermistor element is

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.