Abstract

A methodology based on thermoluminescence dosimetry was developed to check the output of teletherapy units and the given doses. It was applied in a hospital as a part of an extemal quality audit programme. Over a 7 year period the mean ratios of the output doses measured by TLDs calibrated free-in-air to the doses measured at the hospital in a 6 MV X ray and in a 60Co unit were 1.000 +/- 0.024 (n = 86) and 0.997 +/- 0.027 (n=61), respectively. TLDs in capsules were attached to the patient's body or to a phantom to assess entrance, exit and midline doses and transmission. Factors were determined experimentally to relate the doses measured with TLDs in capsules and inside the body. The accuracy in given doses with pelvic and tangential breast fields and assessed via 752 in vivo measurements, was considered to be adequately good, taking into account the limitations of the equipment available in the hospital.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.