Abstract

The primary objective of our study was to develop a thermoelectrical model with both solid and liquid phases to calculate tissue temperature during bipolar coagulation of a posterior spinal artery on the spinal cord. Control of thermal spread caused by coagulation is a concern in spinal surgery. This model utilizes a nonisothermal flow to account for the heat transfer due to the movement of cerebrospinal fluid that is induced by electrical field and temperature gradient. The model is validated by in situ temperature measurements on a porcine spinal cord model. The maximum error for tissue temperature of this model is 12.6%, and the overall average error is 4.2%. The lesional region (>50°C) is identified to be as wide as 5 mm, and thermal dose cumulative equivalent minutes at 43°C (CEM 43) is also calculated with this model. The incorporation of nonisothermal flow has been shown to be crucial in order to accurately predict thermal dose in tissue. The developed model can be further used to establish a guideline for the use of bipolar coagulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.