Abstract

Thermoelectric power sensors can now be used as transfer standards, instead of bolometers, in the microcalorimeter technique. This alternative has the technical advantages to be less sensitive to absolute temperature and not downward frequency limited. At INRiM the high frequency power standards are now based on coaxial thermocouples from dc to 34 GHz. Modified commercial thermocouple mounts in 7 mm and 3.5 mm coaxial line are used to realize the national power standard with an accuracy ranging from 0.03 % to 1 % in the mentioned frequency range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.