Abstract

The radiometry hybrid method, by introducing a lens to the irradiance method with a 900 nm filter radiometer, was adopted to measure the thermodynamic temperature of the aluminum-copper (Al–Cu) eutectic point, aluminum (Al) freezing point, silver (Ag) freezing point, copper (Cu) freezing point and cobalt carbon (Co–C) eutectic point at the National Institute of Metrology, China. By scanning the uniformity of fixed-point blackbodies, the effective diameters were accurately estimated according to the size of the source effect (SSE). The lens transmittance measurement with the closest diameter aperture was adjusted to reduce the uncertainty of SSE. The discrepancy between the thermodynamic temperature and ITS-90 values (T-T 90) of Al–Cu, Al, Ag, Cu and Co–C fixed point blackbodies were concluded as 0.022 °C, 0.014 °C, 0.097 °C, 0.137 °C and 0.317 °C, respectively. The standard uncertainty of the fixed point thermodynamic temperature was estimated to be (0.1 to 0.3) °C (k = 2).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.