Abstract

The Soft X-ray Spectrometer (SXS) instrument (Mitsuda et al., 2010) [1] on Astro-H (Takahashi et al., 2010) [2] will use a 3-stage ADR (Shirron et al., 2012) to cool the microcalorimeter array to 50mK. In the primary operating mode, two stages of the ADR cool the detectors using superfluid helium at ⩽1.20K as the heat sink (Fujimoto et al., 2010). In the secondary mode, which is activated when the liquid helium is depleted, the ADR uses a 4.5K Joule–Thomson cooler as its heat sink. In this mode, all three stages operate together to continuously cool the (empty) helium tank and single-shot cool the detectors. The flight instrument – dewar, ADR, detectors and electronics – were integrated in 2014 and have since undergone extensive performance testing. This paper presents a thermodynamic analysis of the ADR’s operation, including cooling capacity, heat rejection to the heat sinks, and various measures of efficiency.