Abstract

This paper outlines the important features in the calibration of all types of thermocouples and infra-red thermometers using the established methods of calibration. The International Practical Temperature Scale of 1968 is described with reference to the fixed points and methods of interpolation between them. The principle of traceability is described and also the various traceability routes via Absolute, Secondary and Transfer Standards which can be employed. Importance is placed on the provision of written procedures, schedules for routine checking and maintenance of full and accurate records. This work should be the responsibility of a co-ordinating expert. Descriptions are given of temperature sources and ancilliary equipment necessary for establishing a works laboratory which requires traceability to National Standards. The normally attainable levels of uncertainty are given for a number of examples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.