Abstract

Swept-source optical coherence tomography (SS-OCT) demonstrates superior performance in comparison to spectral domain OCT with regard to depth ranging. The main driver of cost for SS-OCT systems is, however, the price of the source. Here we show a low-cost alternative swept source that uses a thermally tuned vertical-cavity surface-emitting laser (VCSEL) at 850 nm. Its center wavelength can be tuned by adjusting the operating temperature through modulation of the injection current. At 2 kHz sweep rate, the depth range of the system was 5 cm, with a sensitivity roll-off of under -3 dB across this range. The system achieved a sensitivity of 97 dB with a sample beam power of 0.3 mW and an axial resolution of 50 µm in air. To demonstrate the system performance in vivo, an eye of a healthy volunteer was measured, and full-eye scans were acquired at 25 and 50 kHz from the cornea to the retina. Based on our results, we believe that this technology can be used as a cost-effective alternative OCT for point-of-care diagnostics.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.