The present work suggests an alternative experimental method in order to not only measure the signal of the deep traps in Al 2 O 3 :C without heating the sample to temperatures greater than 500 °C, but also use this signal for high dose level dosimetry purposes as well. This method consists of photo transfer OSL measurements performed at elevated temperatures using the blue LEDs (470 nm, FWHM 20 nm) housed at commercial Risø TL/OSL systems, after the sample was previously heated up to 500 °C in order to empty its main TL dosimetric trap. The influence of this procedure on specific features such as glow curve shape and sensitivity of the main TL glow peak was also studied.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.