Abstract
A new method to measure low intensity neutrons such as secondary neutrons generated in radiotherapy has been developed. In the proposed method, thermal neutron flux is determined by measuring the conversion electrons from 134mCs generated in a CsI(Tl) scintillator irradiated by neutrons. Because the generated 134mCs is contained in the scintillator as internal sources, the emitted conversion electrons fully stop in the scintillator and counted with high efficiency. The thermal neutron flux in the neutron radiography facility of the Kindai University’s research reactor, UTR-KINKI, was measured by the proposed method. The cadmium difference method was used to separate the contributions of thermal and epi-thermal neutrons. The result was in good agreement with the one obtained by the conventional gold foil activation method. It was concluded that the proposed method is useful for the measurement of thermal neutron flux.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.