Abstract

The indium foil activation technique has been employed to measure thermal neutron fluences ( Φ th) among various locations in the treatment room with a 20×20 cm 2 field size and a 15 and 10 MV X-ray beam. Spatial Φ th are visualized using colored three-dimensional graphical representations; intensities are up to (1.97±0.13)×10 5 and (1.46±0.13)×10 4 n cm −2/Gy-X at isocenter, respectively. The Φ th is found to increase with the X-ray energy of the LINAC and decreases as it moves away from the beam center. However, thermal neutron exposure is not assessed in routine dosimetry planning and radiation assessment of patients since neutron dose contributes <1% of the given therapy dose. However, unlike the accelerated beam limited within the gantry window, photoneutrons are widely spread in the treatment room. Distributions of Φ th were measured in water phantom irradiated with 15 MV X-ray beams. The shielding effect of the maze was also evaluated. The experimentally estimated Φ th along the maze distance was fitted explicate and the tenth-value layer (TVL) was calculated and discussed. Use of a 10 cm-thick polyethylene door placed at the maze was suitable for radiation shielding.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.