Abstract

The applicability of thermal lens spectrometry (TLS) for quantification and routine determination of hexavalent chromium was investigated by using a collinear dual beam thermal lens spectrometer. In aqueous solutions the LOD of 0.1 μg l −1 was obtained for Cr(VI) by using 160 mW laser power. The performance of the technique was verified by the determination of hexavalent chromium in standard reference water samples (NIST SRM 1643a and NIST SRM 1643c) and comparing the results for Cr(VI) in CCA (Cr, Cu, As)-treated timber extracts to concentrations obtained by atomic absorption spectrometry (AAS). Good agreement between the TLS results and reported values for Cr(VI) in SRMs as well as AAS results for Cr(VI) in CCA-treated timbers confirmed that TLS is a reliable and accurate analytical technique applicable for the determination of Cr(VI) in aqueous solutions at concentration levels 0.5–100 μg l −1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.