Abstract
The unique properties of lasers create an enormous potential for specific treatment of chronic ear disease. Despite the widespread acceptance and use of the laser, however, a complete understanding of the time- and space-dependent temperature distribution in otic capsule bone immediately after pulsed laser exposure has not been elucidated. Using a liquid nitrogen-cooled mercury-cadmium telluride infrared detector, the temperature distribution in human cadaveric otic capsule bone was determined immediately after pulsed (100 msec) carbon dioxide laser exposure (0.3 to 4.0 W; 200 microm spot diameter). The time- and space-dependent temperature increases and thermal diffusion were determined as a function of the laser power density and were found to vary linearly.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.