Abstract

This paper contains descriptions of the construction and use over the temperature range -27 °C to 570 °C of a Merritt-Saunders (optical interferometric) linear thermal expansion apparatus. Measurements of thermal expansion are reported for platinum and for two platinum-rhodium alloys (nominally 12 wt% Rh and 20 wt% Rh). Detailed analyses are given of the measurement uncertainties involved in the experiment and of the representation of the data by polynomials in the sample temperatures. The data show precision at the 1-ppm level and good agreement with results already published.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.