Abstract

To determine the effect of covering infant incubators on incubator wall and air temperatures, as well as on infant temperatures. A within-subject ABA design, in which blankets covering the incubators were removed for a 30-minute period and then replaced. A neonatal intensive-care unit. Eight medically stable infants (gestational age, 28-33 weeks; birth weight, 913-1,947 g; and postnatal age, 2-39 days). Incubator wall and air temperatures as well as infant temperatures were measured during three study conditions: incubators covered (30 minutes), uncovered (60 minutes), and re-covered (30 minutes). Incubator air and wall temperature; infant temperature. All incubator walls decreased in temperature after being uncovered; the decrease ranged from 0.6-2.2 degrees C. Although infants maintained relatively stable body temperatures during the uncovered period, the energy cost to their thermoregulatory efforts is unknown.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.