Abstract
<para xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> An ultrasonic density-measurement cell is experimentally characterized as a function of temperature. The measurement of propagation velocity and density of several liquids is performed in the 15 <formula formulatype="inline"><tex>$^{\circ}\hbox{C}$</tex></formula>–40 <formula formulatype="inline"><tex> $^{\circ}\hbox{C}$</tex></formula> temperature range. Results are compared to the tabulated values in the case of distilled water, showing an accuracy of 0.07% for the propagation velocity. The cell was tested with distilled water, alcohol, and homogenized milk, and density values are compared to those obtained with a pycnometer, showing 0.2% accuracy in density measurement for stabilized temperature and 0.4% accuracy under thermal gradient conditions. </para>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.