Abstract

Scanning electron microscopy (SEM) sometimes induce defects on samples during imaging. We study in this article the thermal effects of SEM views on the Cu/BCB interface, in focused ion beam (FIB) cross sections. Electrons can lead to local thermal power dissipation due to their deceleration and create delamination. A TEM lamella sample preparation method was also found to avoid this kind of delamination. In this case, the thermal dissipated power could be reduced due to the fact that a big part of the incoming electrons could go through the sample without interactions. Several thermal simulations (2D/3D) were carried out to estimate the field of temperatures under electron beam and to explain the Cu/BCB interface delamination.