Abstract

A study on the thermal analysis of a new tin fixed-point open cell within a new three-zone furnace was carried out. The stability at the setting temperatures of liquid-solid coexisting together with some degree Celsius lower and higher than the tin fixed point; and the axial uniformity of furnace while tin is still in solid phase were investigated and analyzed. The impurities effect on the depression in temperature was investigated in terms of ΔT (Tobserved-T1/F=0) and the inverse of the melted fraction (1/F) relationship during the melting and the following freezing realizations at various temperature settings of furnace. These thermal analysis results were also compared with those estimated by the CCT-WG1 recommended SIE (sum of individual estimates) method, which leads to a temperature correction along with a corresponding uncertainty through the individual impurity content detected by GDMS (glow discharge mass spectrometry).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call