Abstract
The Spectra Optia apheresis system has only recently been approved by the Food and Drug Administration (FDA) for therapeutic white blood cell (WBC) depletions and is not yet approved for platelet depletions. Prior to FDA-approval of the WBC depletion protocol, when our available COBE Spectra apheresis systems were out of service, we successfully performed WBC depletion using a modified Spectra Optia apheresis system Continuous Mononuclear Cell (CMNC) protocol. Using this modified Spectra Optia CMNC protocol, we created institutional protocols for WBC and platelet depletions. We performed 10 WBC depletions in 9 patients and 2 platelet depletions in 2 patients. We compared pre- and post-procedure WBC, platelet count, and hemoglobin to the same data from patients previously treated on the COBE Spectra and found no difference in % WBC and platelet reduction. We also found no significant difference in post-procedural hematocrit decline. Additionally, adverse reactions were not increased. Therefore, we conclude that the Spectra Optia CMNC protocol can be successfully modified for effective WBC and platelet depletions without increase in adverse reactions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.