A reliable liquid chromatography-tandem mass spectrometric(LC-MS/MS) method was developed and validated for the simultaneous determination of vancomycin and voriconazole in human plasma. The analytes and internal standard 10-hydroxycarbazepine were separated at a flow rate of 0.9 mL/min using a Zorbax SB-C18 column (50 mm×4.6 mm, 2.7 μm). Positive ion electrospray ionization was used to detect vancomycin, voriconazole and intrenal standard(IS) 10-hydroxycarbazepine followed by multiple reaction monitoring(MRM) of the transition at m/z 725.5→144.2, 350.3→281.0 and 253.1→208.0, respectively. The total run time for both vancomycin and voriconazole samples was 5 min; 0.30 μg/mL was the lower limit of quantification. The precision of intraday and interday was no more than 12.4%. The method was successfully and resoundingly applied in therapeutic drug monitoring of 156 patients treated with vancomycin and voriconazole.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.