Abstract
A simple analytic model is presented which is used to calculate the impulse response of a computed tomography system for parametric variations of the geometry. The model is useful for fan beam as well as parallel beam geometries. Results show that wide-aperture detector systems have intrinsic resolution limitations that are not mitigated by oversampling. Furthermore, the optimum focal spot size of the X-ray source is found to be dependent on the detector geometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.