Abstract

The signal-to-noise ratios for the path-stabilized, the active-heterodyne, the Sag, and the Fabry-Perot interferometers are calculated for cases in which dynamic surface displacements are detected. Expressions for the minimum-detectable displacements are given for the sensitivity limit of each interferometer. It is found that the ultimate sensitivities of the interferometers considered are nearly equal for similar conditions of use. Sensitivity values are shown to be about 10−15 m Hz−1/2 for systems that use typical laboratory instruments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.