Abstract

The author shows theoretically that when a magnetised blood bolus enters a CW NMR excitor coil of length Le at resonance and the signal from the T2-decaying, precessing transverse magnetisation of the flowing blood spins is subsequently detected by a detector coil of length L separated from the excitor coil by a distance Delta l, then by recording CW NMR signals at three positions such as Delta l=0, 0.5 and 1.0 cm one can eliminate the static tissue signal and measure non-invasively the steady component V0 as well as the total vessel cross section, beta accurately. The time dependent part of the CW NMR signal which depends on Vpulse(t), is also dependent on V0 non-linearly unless both L and Le are greater than 50 cm and Delta l is zero. Finally, methods of obtaining true Vpulse(t) from the CW NMR signal after applying proper correction due to the steady flow are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.