Abstract
Effects of flow oscillations on spectrum of Ultrasound Doppler Velocimetry (UDV) signals were investigated theoretically and numerically. A laminar pipe flow with a superimposed oscillating component was considered. Negative impact of oscillation on the ultrasound signal hence on the flow images was observed in the form of spreading of spectral ultrasound signal energy around mean component, leading to image artifacts. Both analytical and numerical results revealed the strong effect of a group of parameters including Doppler frequency, flow oscillation amplitude and frequency. Exceeding a particular value of the group, 1.45, resulted in artifacts in the flow images. Revealing the mechanisms involved in the deteriorations associated with the flow oscillations is potentially useful in UDV studies involving random flow fluctuations such as turbulence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.