Abstract

A theoretical approach is presented improving the knowledge on real time fast neutron dosimetry in intense fields. The detector is a reverse biased silicon junction operating at room temperature. The signal readout is a current measurement for which the leakage current of the sensor sets the detector threshold in the range of 1 Sv s −1 dose rates. The dosemeter energy response as a function of shielding, polyethylene neutron converters and diode characteristics has been computed to provide a tissue equivalent neutron monitor which is as insensitive as possible to the gamma photon background. Two Monte Carlo codes, one for the fast neutrons (SEMIC) and for the photon background (MCGET), are used to calculate the current response of the device irradiated by monoenergetic radiations (n or γ) as well as neutron sources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.