Abstract

The response of some SSNTDs was calculated theoretically taking into consideration the major parameters that affect them. These parameters are related to the behavior of alpha emitters, the detectors types and the used diffusion chambers. The obtained results showed that the response of the filtered CR-39 detector is about 5 times higher than that of the LR-115 detector and more than 6 times higher in case of bare track detectors. This calculation was used to determine the optimum diffusion chamber dimensions with experimental verification for different diffusion chambers. The cup of 3 cm radius and 5 cm height gives a good agreement with the theoretical value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.