Abstract

SummaryDetailed investigating into the effects of spherical and chromatic aberrations, diffraction and the probe current allows the more general formulae for the optimized aperture and the minimum probe radius in low‐voltage scanning electron microscopes (LVSEMs) to be derived using both wave optics and geometric optics. The probe sizes for a diversity of electron sources in LVSEMs have been estimated, which may be useful for practical applications. The computed results indicate the possibility of achieving 1·5–2·0‐nm resolution at low voltages.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.