Abstract

Temperature distribution in tissues behind ribs produced by weakly focused ultrasound had been calculated using Pennes bioheat equation and the validity of the theoretical model was experimentally confirmed in vitro using porcine live. We found that the position of the maximum temperature in tissue is strongly influenced by the distance between the transducer and ribs, while the gap between ribs is the determining factor for the maximum achievable temperature. Within the focal length, when the distance between the transducer and ribs increases, the maximum temperature increases and its position shifts away from the transducer. The rib width has little effect on the position of the maximum temperature but affects the achievable peak temperature. Our results provide useful information for treating liver cancers using ultrasound induced hyperthermia. (E-mail: dzk@psu.edu)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.