Abstract
This study examines the validity of labeled parameters of aphakia-correcting soft contact lenses. The Nikon Projection Lensometer (NPL) and the Hydro-Vue Soft Lens Analyzer (HSLA) were used by two independent inspectors to measure 51 aphakia-correcting lenses. The contact lens power, diameter, and base curve were measured by each of the inspectors and compared to the labeled parameters. All parameter relations that were analyzed revealed that measurements made by the inspectors were not statistically different (p greater than 0.05) from each other and that their combined measurements were not statistically different (p greater than 0.05) from the labeled values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: American journal of optometry and physiological optics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.