Abstract

IntroductionThe distance between a coronal reference point and the major apical foramen is important for working length determination. The aim of this in vitro study was to determine the accuracy of root canal length measurements performed with cone-beam computed tomographic (CBCT) scans using a gold standard. MethodsA total of 162 teeth (198 root canals) in 16 dry human dentulous mandibles were scanned using a 3DX-Accuitomo CBCT scanner (Morita 3DX; J Morita Mfg Corp, Kyoto, Japan). The root canal length was measured with CBCT data. All teeth were extracted atraumatically and endodontically accessed; the root canal length was measured blindly using a #10 K-file (Dentsply Maillefer, Ballaigues, Switzerland) and served as the gold standard. ResultsThe mean absolute difference of the CBCT-based root canal length from the gold standard was 0.46 mm (95% confidence interval, 0.41–0.50 mm). Only in 9 of 198 (4.5%) roots did the difference between the CBCT-based root canal length and the gold standard exceed 1 mm. ConclusionsCBCT-based root canal length measurements are accurate and reliable when compared with a gold standard.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.