Abstract

Transvaginal ultrasound probes (TVUS) are used for several gynecological procedures. These need to be disinfected between patient use. In the current study we examine whether UVC delivered using light emitting diodes for 90 seconds can provide sufficient disinfection efficacy. A new UVC device that delivers UVC radiation at 265nm-275nm for 90 seconds was used. TVUS probes were swabbed before and after use in an in vitro fertilization clinic. Microbes on the swabs were cultured and identified. In addition, the ability of the UVC device to provided repeated high-level disinfection was analysed by deliberately contaminating probes with spores of Bacillus subtilis and then performing the UVC disinfection and bacterial culture. 50% of probes were contaminated with bacteria, most commonly Bacillus sp., directly after in vivo use. Whereas 97% were sterile after UVC disinfection for 90 seconds. The UVC treatment resulted in no growth of B. subtilis spores after each of five repeated contaminations with 5-9 x 107 spores on the probes. This study has found that UVC delivered via light emitting diodes for only 90 seconds can produce high level disinfection of transvaginal probes.