Abstract
We compared the performance characteristics of a commercial dilute Russell's viper venom (DRVV)-based APC resistance assay (Gradipore PC Impedance Test) to a routinely utilized commercial APTT based assay (Coatest APC Resistance Assay). The DRVV based assay offers improved sensitivity and specificity for the factor V Leiden mutation. However, the routine use of both assays provides optimum reliability for diagnosis of genetic APC resistance. Our results suggest that when both tests are either positive or negative, DNA analysis is unnecessary. Interference by lupus anticoagulants is dramatically minimized by the phospholipid rich DRVV reagent used in the assay and it is insensitive to high factor VIII activity. Additionally, discrepant functional assay results allow identification of patients who may have an acquired APC resistant phenotype.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.