Abstract

A simple visual and tristimulus colorimetric method (threedimensional system CIE-L*a*b*) for the determination of trace amounts of hydrogen cyanide in air has been developed. The method is based on the suction of hydrogen cyanide through a chlorinating cartridge where cyanogen chloride is formed, which is further driven to an indicator disc made of a modified cotton fabric. This indicator disc is placed into an adapter. Prior to analysis, the disc is saturated with a chromogenic reagent, a solution of 5,5-dimethyl-1,3-cyclohexanedione (dimedone) and 4-benzylpyridine in ethanol. In the presence of hydrogen cyanide (cyanogen chloride), a pink coloration emerges on the indicator disc, the intensity of which is evaluated either visually or by use of a tristimulus colorimeter. The detection limit is 0.1 mg m-3. The method is mainly suitable for mobile field analyses. It was applied for the CHP-5 chemical agent detector introduced into the equipment of the Czech Army corps.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.