Abstract

Undesirable temperature gradients in a NMR sample tube are usually generated by an inappropriate temperature regulation system. We have shown that such convection effects can greatly distort the measurement of translational self-diffusion coefficients. The use of sample spinning helps to minimize such undesirable effects by disruption of convection fluxes due to resulting Coriolis forces that have a strongly stabilizing effect on the conducting state of the system (J. Lounila et al., J. Magn. Reson. A 118, 50 (1996)). This simple trick allows the accurate measurement of diffusion coefficients for a wide range of temperatures and solvents without the need for a convection-compensated NMR pulse sequences or more sophisticated temperature control units. Experimental data obtained for some target compounds dissolved in several common deuterated solvents at different temperatures are reported and discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.