Abstract

Exit waves can be reconstructed from through focus series of HREM images or by off-axis holography [1]. We have applied the through focus method to reconstruct exit waves, using algorithms developed by Van Dyck and Coene [2]. Electron microscopy was performed with a Philips CM30ST electron microscope with a field emission gun operated at 300 kV. The high resolution images were recorded using a Tietz software package and a 1024x1024 pixel Photometrix CCD camera having a dynamic range of 12 bits. The reconstructions were done using 15-20 images with focus increments of 5.2 nm. The resulting exit waves were corrected posteriorly for the three fold astigmatism.The exit wave is complex; consequently it contains phase and amplitude. Since in the very thin regions the specimen acts as a thin phase object, such a thin area will show little contrast, an example of which is shown in Figure 1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.