Abstract

This study evaluated the precision of electronic working length by microcomputed tomography using two electronic apex locators (EALs). Twenty single-rooted permanent teeth without caries or restorations were selected as the subject teeth. The positions of the minor apical constriction (AC) and major apical foramen (AF) were measured by electronic root canal length, and microcomputed tomography was performed with the file inserted and fixed in the root canal. All teeth were measured individually and independently by two operators. The Mann-Whitney U-test was used to statistically test the AC and AF values using two EALs; P < 0.05 was defined as statistically significant. This was 65.0% within 1.5 mm in the case of two EALs on AC. This was more than 90.0% within 1.0 mm in the case of two EALs on AF. Comparison of the differences between the respective AC and AF of the measurements obtained using the two EALs revealed no significant difference. The two EALs are devices that can greatly improve the accuracy of WL control.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.