Abstract
The problem of filtering artifact from QRST complexes of ECG records has been approached by the use of a digital, time-varying filter. The optimum form of this filter assuming uncorrelated, zero-mean, additive noise is presented and shown to be equivalent to estimating the coefficients of an intrinsic component expansion of the complex. Results are reported showing that QRST complexes in limb leads of normals may be accurately represented by only eight terms of the intrinsic component expansion. A suboptimum filter is presented and results reported which show that an order of magnitude reduction of the 0.05 to 100 Hz artifact energy in the filtered complex can be achieved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.