Abstract

To present the functional and radiographic outcome 1 and 6 years after application of a new intramedullary fixation device for proximal humerus fractures. Retrospective case series. Level II orthopaedic surgery hospital. Twenty-six consecutive patients (average age 68.9 years) with 2-, 3- and 4-part fractures of the proximal humerus were operated at a single institution. Follow-up was performed after 1 year (26 patients) and 6 years (16 patients). All patients were treated with closed reduction and intramedullary helix wires. The Constant-Murley score and the University of California Los Angeles (UCLA) score. Clinical complications and radiological posttraumatic arthritis were recorded. The average Constant-Murley score was 70.3 (points) and 70.7 after 1 and 6 years, respectively; the average UCLA score was 27.2 and 31.5 after 1 and 6 years, respectively. Major complications were 4 revisions for 3 secondary fragment displacements and 1 nonunion with partial avascular osteonecrosis in the first postoperative year. Complications were found predominantly in 4-part fractures (3/5, 60%). There were no further complications or progressive posttraumatic arthritis up to 6 years following surgery. The helix wire is well suited for displaced or unstable 2- and 3-part proximal humerus fractures. Adequate functional outcome, a low number of implant displacements, a low number of application morbidity, and infrequent implant removals were recorded. The use of this device is not recommended for 4-part fractures.